The ATAGO Model Polax-2L Semiautomatic Polarimeter
is the ideal instrument for measuring the optical
activity of both inorganic and organic compounds.
A compound is considered to be optically active if
linearly polarized light is rotated when passing through
it.
Easy to Use. Operation consists of filling
one of the sample tubes with the sample solution and
placing the tube on the polarimeter stage. When viewed
through the eyepiece, the left and right visual fields
will have different brightness levels. The rotation
switches are adjusted until fields are of equal brightness.
At this point the correct reading can be viewed on
the large LED display. The unit calibrates with water
- a zero indicator lets the user know the unit is
calibrated.
Semiautomatic. The rotation switches have both
fast and slow push-button modes. The fast mode allows
the Polax-2L to be quickly adjusted to an approximate
reading by continuously rotating the analyzer. The
slow mode allows the Polax-2L to be accurately adjusted
by rotating the analyzer 0.05° (0.1°Z) at
a time.
Digital Accuracy. High resolution and accuracy
provide substantial precision for the testing of optically
active solutions. The digital display and push-button
operation both enhance overall precision - no dials
to turn, no vernier scales to read.
Electronic Temperature Measurement. Model Polax-2L
features a temperature measurement range of 0.0-50.0°C
for the sample chamber; it's displayed on the LED
display with the push of a button.
LED Light Source. The long lasting LED light
sources with 589nm interference filter is intended
to last the life of the instrument. Model Polax-2L
automatically turns off when the unit is left inactive
for a period of more than 5 minutes.
|
|
|
Two Measuring Scales. For flexibility, Model Polax-2L comes equipped
with both Angle of Rotation and the International Sugar Scale.
Wide Range of Industries and Applications. Sugar refineries, University
research, Flavors, Fragrances, Essential Oils, Identification and Characterization
of chemicals, Sugar concentration and Purity in food products.
Sample Tubes. Two end-fill sample tubes (100mm and 200mm) are included
with the unit. The 200mm tube permits higher accuracy when used with transparent
liquids, while the 100mm tube should be used with cloudy samples containing
impurities. The tubes also contain a bubble trap which prevents errors caused
by air bubbles in the sample. When only a small amount of sample is available
for measurement, an optional micro sample tube (100mm, 1.0-1.5ml) is available.
|
Measuring System |
Semiautomatic
Polarimetry |
Measuring Range |
Angle of Rotation: |
+180.00° to 0.00° to -179.95° |
Resolution |
International Sugar Scale: |
+130.0°Z to 0.0°Z to -130°Z |
Accuracy |
Angle of Rotation: |
±0.05° |
International Sugar Scale: |
±0.1°Z |
Ambient temperature |
5-40°C |
ATC - Automatic Temperature Compensation |
NO |
Measuring Temperature |
0.0 - 50.0°C |
Relative Humidity |
<90% |
Altitude |
<5000m Above Sea
Level |
Calibration Liquid/ Medium |
Distilled Water |
Display |
LED |
Polarizing Light Source |
LED with Interference
Filter (589nm) |
Power Supply |
100 to 240V AC, 50/60
Hz |
Power Consumption |
20W |
Dimensions |
22 x 43 x 30 (L x W
x H) cm |
Weight |
12kg |
Supplied With:
Power Cord (1 ea)
100mm Sample Tube (1 ea)
200mm Sample Tube (1 ea) |
Fuse (1 ea)
Instruction Manual (1 ea)
|
|
|
Right Rotation:
When
the right semicircular field is brighter, press the RIGHT ROTATE button
until the two fields are of equal brightness. |
|
Left Rotation:
When
the left semicircular field is brighter, press the LEFT ROTATE button until
the two fields are of equal brightness. |
|
|
|
Weight:
30.00 Pounds |
Dimensions:
1.00(L) x
1.00(W) x
1.00(H) Inches |
Weight Disclaimer: *Weight may vary based on configuration. |
|
|
|
|
|